GDYD-55D AC/DC Hipot Test Set

GDYD-55D AC/DC Hipot Test Set

Brief Description:

GDYD series AC/DC Hipot Test Set is composed of manual control unit and dry testing transformer (HV unit) which output AC voltage smoothly. When adding a HV diode, it can output DC voltage. Meanwhile, it has many monitoring and protection functions. 


Product Detail

Product Tags

Features

• Digital display meters mounted on the panel.

• Monitoring HV side voltage, HV side current, LV side current and indicator of zero, power, working start, timing.

• Over-current protection, zero-starting protection, sound and light alarm.

• With new type time relay, time rang is wider (1S ~ 99H).

• Using the latest current relay, more accurate and reliable.

• Light weight, small size, easy to move.

Specifications
  1. A .Control unit specifications

• Power: AC220V, 50Hz

• Capacity: 5kVA

• Output voltage: AC 0-250V

• Rated output current: 25A

• Rated current rectified range: 0-25A

• HV voltmeter range: 0-70kV

   Accuracy: 3.0% of the reading±1digit (F.S)

• HV ammeter range: 0-100mA  

Accuracy: 3.0% of the reading±1 digit (F.S)

• LV ammeter range: 0-25A

Accuracy: 2.0% of the reading±1 digit (F.S)

• Timing range: 1s-99h

• Weight: 25kg

• Dimension: 470mm (L)*370mm (W)*240mm (H)

 

B. Testing transformer (HV unit) specifications

• LV side voltage (Input voltage): 0-200V

• LV side current (Input current): AC 0-25A

• HV side voltage (Output voltage): AC 0-50kV, DC 0-70kV

• HV side current (Output current): AC 100mA, DC71.4mA

• Capacity: 5kVA

• HV side/LV side turns ratio: 250:1

• Measuring voltage: AC 0-100V

  Accuracy: 1.5% of the reading±1 digit (F.S)

• No-load current: <10%

• Impedance voltage: <10%

• Weight: 35kg

• Dimension: 280mm(L)*320mm(W)*310mm(H)

 

C. Working conditions

• Working place: No gas, steam, chemical dust and explosive and corrosive medium which will influence insulation.

• Environment temperature: -20--50℃

• Environment humidity: ≤85% RH (no condensing)